Health Products

Allura Systems

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Allura Systems
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
Allura Xper FD10 (Product Of Alluraclarity Family)All lots.722026
Allura Xper FD20/20 OR Table (Product Of Alluraclarity Family)All lots.722039
Allura Xper FD10/10 (Product Of Alluraclarity Family)All lots.722027
Allura Xper FD20/15 OR TableAll lots.722 059
Allura Xper FD20 (Product Of Alluraclarity Family)All lots.722028
Allura Xper FD20/10 (Product Of Alluraclarity Family)All lots.722029
Allura Xper FD20/20 (Product Of Alluraclarity Family)All lots.722038
Allura Xper FD20/15All lots.722058
Allura Xper FD10 OR Table (Product Of Alluraclarity Family)All lots.722033
Allura Xper FD20 OR Table (Product Of Alluraclarity Family)All lots.722035

Issue

Philips has identified a software issue in the internal communication process between the system software and the x-ray generator firmware which may lead to temporary loss of imaging (x-ray) functionality. This issue may be triggered under either of the following situations: Situation 1 - Pedal tap: When the foot switch is pressed and released rapidly without generating x-rays (due to a short duration of contact), the system may enter a state where x-ray generation is inhibited. When this occurs, the system does not show any message to the user indicating that x-ray is not possible. A warm restart may resolve the situation in some cases. A cold restart will always resolve the situation. This issue can lead to a delay in therapy. Situation 2 - Phase fault: Allura systems are designed to automatically recover from phase faults (a phase fault is an abnormal condition in which one or more phase voltages drop to (near) zero). When a phase fault occurs, the x-ray generator software initiates an automatic recovery process to restore fully system functionality, during which imaging (x-ray) functionality is not available for approximately 5 seconds. When the identified software issue occurs during this recovery, communication with the system software is lost, resulting in the system not recognizing that the generator has recovered and is ready for operation. In this situation, the system shows the user message "generator is busy starting up, no x-ray possible". If the phase fault is detected by the generator during acquisition, the system also shows the message "run aborted: tube problem". to restore system functionality a cold restart of the system is required. These issues affect the Allura Xper systems with software release 8.2 or higher, equipped with a Certeray generator. Recall start date: July 7, 2025