Novum IQ LVP

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Novum IQ LVP
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue
Novum IQ LVPAll lots.40700BAX

Issue

Baxter is issuing an urgent medical device correction for the Novum IQ Large Volume Pump (LVP) due to the following two issues: there is a potential for underinfusion when transitioning from a lower to a higher flow rate (e.g., rate change or bolus). Specifically, the risk occurs when the second flow rate is more than double the first flow rate. Baxter has identified an increase in customer reports of over and underinfusion that may be due to set misloading. Failure to properly load the tubing into the pump channel may result in the pump infusing at a rate higher or lower than programmed. Recall start date: July 17, 2025