Health Products

Centrella Smart+ Bed

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Centrella Smart+ Bed
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Centrella Smart+ BedMore than 10 numbers, contact manufacturer.P7900B100205
Centrella Smart+ BedMore than 10 numbers, contact manufacturer.P7900B100021
Centrella Smart+ BedMore than 10 numbers, contact manufacturer.P7900B100020
Centrella Smart+ BedMore than 10 numbers, contact manufacturer.P7900B100019
Centrella Smart+ BedMore than 10 numbers, contact manufacturer.P7900B100223
Centrella Smart+ BedMore than 10 numbers, contact manufacturer.P7900B100224
Centrella Smart+ BedMore than 10 numbers, contact manufacturer.P7900B100231

Issue

Baxter Corporation is issuing an urgent medical device correction for Centrella Smart+ beds with system software version 1.27.000 due to the potential for an interface issue with the Rauland Responder 5 nurse call system. The bed exit system will alert at the bed but may fail to send a remote alert through the nurse call system if a remote alert was previously sent and cancelled at the in-room nurse call wall unit. If the bed exit alert fails to notify the caregiver at the assigned remote location (e.g., nurse's station) the staff may be unaware the patient exited the bed if they are unable to hear the local bed alert. As a result, a patient may experience a fall resulting in a critical musculoskeletal injury, including head trauma. To date, Baxter has not received any reports of patient injury related to this issue. Recall start date: September 19, 2024

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime