Health Products

VITROS® Chemistry Products System - LAC Slides (Lactate)

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: VITROS® Chemistry Products System - LAC Slides (Lactate)
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalog number
VITROS® Chemistry Products System - LAC Slides (Lactate)More than 10 numbers, contact manufacturer.815 0112
VITROS® Chemistry Products System - LAC Slides (Lactate)More than 10 numbers, contact manufacturer.843 3880

Issue

QuidelOrtho investigated customer complaints related to the inability to successfully calibrate VITROS® LAC slides on VITROS® XT 3400 systems and VITROS® XT 7600 systems. To resolve this issue, QuidelOrtho has adjusted the assay parameter used by the reflectometer to process VITROS® LAC slides. This parameter adjustment will be implemented beginning in add DRV 6339 (and above). QuidelOrtho advises affected customers to load add DRV 6339 (or above) upon receipt. Recall start date: August 5, 2025