· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Unicel DxH 900 Coulter Cellular Analysis System | More than 10 numbers, contact manufacturer. | C11478 |
| Unicel DxH 600 Coulter Cellular Analysis System | BB49881 | B23858 |
| Unicel DxH 800 Coulter Cellular Analysis System | AU10078 | 629029 |
| Unicel DxH 800 Coulter Cellular Analysis System | AW43276 | 629029 |
| Unicel DxH 800 Coulter Cellular Analysis System | AZ46681 | 629029 |
| Unicel DxH 800 Coulter Cellular Analysis System | AU10078 | B24802 |
| Unicel DxH 800 Coulter Cellular Analysis System | AW43276 | B24802 |
| Unicel DxH 800 Coulter Cellular Analysis System | AZ46681 | B24802 |
| Unicel DxH 690T Coulter Cellular Analysis System | BJ12047 | C34520 |
| Unicel DxH 690T Coulter Cellular Analysis System | BJ07013 | C34520 |
| Unicel DxH 690T Coulter Cellular Analysis System | BJ12046 | C34520 |
| Unicel DxH 690T Coulter Cellular Analysis System | BH24076 | C34520 |
Beckman Coulter has identified a potential for erroneously high hemoglobin (Hgb) results on samples with elevated white blood cell (WBC) counts. This occurs on Unicel DxH series instruments configured with the led Hgb photometer. Falsely elevated Hgb results may observed in specimens with WBC counts > 30 x 10³ cells/μl. Recall start date: August 6, 2025