Health Products

Total Knee Arthroplasty (TKA) and Total Hip Arthroplasty (THA)

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Total Knee Arthroplasty (TKA) and Total Hip Arthroplasty (THA)
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
TKA 2.0 + TKA1.0.1+ THA 4.1. + PKA 3.0.2 + MGO 1.2.2Not applicable.700001590414
TKA 2.0 + TKA1.0.1+ THA 4.0.0.1 + PKA 3.0.2 + MGO 1.2.2Not applicable.700001590415

Issue

Stryker discovered specific to the applications listed above, an increase in the software error #3 (se3) error code when a Mako® system shutdown or a Mako® system restart is not performed prior to switching between applications (i.e. TKA to THA). Recall start date: August 18, 2025