Total Knee Arthroplasty (TKA) and Total Hip Arthroplasty (THA) Aug 28, 2025 · Published by Health Canada · Official notice
Share
Alert type : Health product recall
Category : Health products - Medical devices - General hospital and personal use
Product : Total Knee Arthroplasty (TKA) and Total Hip Arthroplasty (THA)
Issue : Medical devices - Performance
What to do : Contact the manufacturer if you require additional information.
Audience : Healthcare
Recall class : Type II
We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.
Get health product recall alerts CA$3.99/mo · cancel anytime
Affected products Lot or serial number Model or catalogue number TKA 2.0 + TKA1.0.1+ THA 4.1. + PKA 3.0.2 + MGO 1.2.2 Not applicable. 700001590414 TKA 2.0 + TKA1.0.1+ THA 4.0.0.1 + PKA 3.0.2 + MGO 1.2.2 Not applicable. 700001590415
Stryker discovered specific to the applications listed above, an increase in the software error #3 (se3) error code when a Mako® system shutdown or a Mako® system restart is not performed prior to switching between applications (i.e. TKA to THA). Recall start date: August 18, 2025