Getinge Washer/Disinfector

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Getinge Washer/Disinfector
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Getinge Washer/DisinfectorW5003731688-5

Issue

The issue found can cause the device to overheat during operation if the circulation pressure calibration was not performed or been performed incorrect and the low-pressure alarm is not triggered. The scope of this action is the electrical heated 88-5 washer disinfectors. Justification: the construction on steam heated devices is different and the heating elements cannot be overheated. This field action is initiated to verify that the circulation pressure in standby mode reads 0.0 kPa. If the value deviates from 0.0 kPa, the device must be recalibrated by a qualified technician to ensure safe continued clinical use. Recall start date: August 21, 2025