· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Equinoxe® Reverse Shoulder System - Humeral Liner | More than 10 numbers, contact manufacturer. | 320-38-00 |
| Equinoxe® Reverse Shoulder System - Humeral Liner | More than 10 numbers, contact manufacturer. | 320-38-03 |
| Equinoxe® Reverse Shoulder System - Humeral Liner | More than 10 numbers, contact manufacturer. | 320-42-00 |
| Equinoxe® Reverse Shoulder System - Humeral Liner | More than 10 numbers, contact manufacturer. | 320-42-03 |
Exactech is recalling Equinoxe® humeral liners due to non-conformance identified during inspection at Exactech's Sarasota facility for liners as they were found to have spherical articular surface positioned out of specification relative to the engineering drawing. A manufacturing non-conformance has been identified during inspection of Equinoxe® humeral liners. Specifically, the spherical articular surface of certain liners may be positioned outside of specified tolerances, which could adversely affect joint alignment and overall implant performance. Recall start date: August 15, 2025