Health Products

MOSAIQ

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: MOSAIQ
Issue: Medical devices - Performance
What to do: A field safety notice will be sent. A field safety modication will follow.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
MOSAIQ3.2.13.2.1.0 (01)07340201500071(10)3.2.1.0
MOSAIQ3.2.13.2.1.1 (01)07340201500071(10)3.2.1.1
MOSAIQ3.2.13.2.1.2 (01)07340201500071(10)3.2.1.2
MOSAIQ3.2.13.2.1.3 (01)07340201500071(10)3.2.1.3
MOSAIQ3.2.13.2.1.4 (01)07340201500071(10)3.2.1.4
MOSAIQ3.1.33.1.3.0 (01)07340201500026(10)3.1.3.
MOSAIQ3.1.33.1.3.1 (01)07340201500026(10)3.1.3.1
MOSAIQ3.1.33.1.3.2 (01)07340201500026(10)3.1.3.2
MOSAIQ3.1.33.1.3.3 (01)07340201500026(10)3.1.3.3
MOSAIQ3.1.33.1.3.4 (01)07340201500026(10)3.1.3.4
MOSAIQ3.2.23.2.2.0 (01)07340201500071(10)3.2.2.0
MOSAIQ3.2.23.2.2.1 (01)07340201500071(10)3.2.2.1
MOSAIQ3.2.3.03.2.3.0 (01)07340201500071(10)3.2.3.0

Issue

When appending a care plan with an ad hoc waved order set, if an order does not fall on a care plan cycle day, it is not appended. If the care plan is not appended, it may result in one or more doses of the medication(s) not being administered. Recall start date: August 29, 2025