myAirvo™ 2 and Airvo™ 2 Humidifier

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: myAirvo™ 2 and Airvo™ 2 Humidifier
Issue: Medical devices - Performance
What to do: Customers will be notified with instructions to follow.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
myAirvo™ 2 Humidifier09420012422248PT100US
Airvo™ 2 Humidifier09420012422347PT101US

Issue

This recall is to inform customers of an update to the disinfection kit user manual for maintaining the Airvo™ and myAirvo™ 2 device throughout its expected service life. The Airvo™ 2 and myAirvo™ 2 device has an audible alert that sounds for at least 120 seconds when the unit is unintentionally disconnected from a power source. The disinfection kit user manual has been updated to include a regular test of this alert, to be carried out in between patient uses. This update has been made due to reports of the audible alert sounding for less than 120 seconds, which may delay user awareness to reconnect power. There have been no events with serious injuries or death related to this issue. Recall start date: August 27, 2025