Health ProductsType I recall

HeartSine® Pad-Pak™

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: HeartSine® Pad-Pak™
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
HeartSine® Pad-Pak™More than 10 numbers, contact manufacturer.PAD-PAK-01
HeartSine® Pediatric-Pak™More than 10 numbers, contact manufacturer.PAD-PAK-02
HeartSine® Pad-Pak-07 Battery & Defibrillation PadsMore than 10 numbers, contact manufacturer.PAD-PAK-07

Issue

Post-market surveillance has revealed that the Pad-Paks are not always inserted properly into the HeartSine Samaritan® PAD devices as outlined in the instructions for use (IFU), which can create failure during device use. Upon investigation, the potential causes of the improper insertion of Pad-Paks include use error and bent locator pins, which may occur during the manufacturing process. In the event the device is unable to complete connection, the device will repeatedly prompt "apply pads to patient's bare chest". In some cases, the AED device may fail to power on entirely. Recall start date: August 22, 2025