Health Products

TactiFlex™ Ablation Catheter, Sensor Enabled™

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: TactiFlex™ Ablation Catheter, Sensor Enabled™
Issue: Medical devices - Labelling and packaging, Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
TactiFlex™ Ablation Catheter, Sensor Enabled™More than 10 numbers, contact manufacturer.A-TFSE-DF
TactiFlex™ Ablation Catheter, Sensor Enabled™More than 10 numbers, contact manufacturer.A-TFSE-FJ
TactiFlex™ Ablation Catheter, Sensor Enabled™More than 10 numbers, contact manufacturer.A-TFSE-D
TactiFlex™ Ablation Catheter, Sensor Enabled™More than 10 numbers, contact manufacturer.A-TFSE-F
TactiFlex™ Ablation Catheter, Sensor Enabled™More than 10 numbers, contact manufacturer.A-TFSE-DD

Issue

Abbott is issuing this recall letter related to the TactiFlex™ Ablation Catheter, Sensor Enabled™. As part of their postmarket surveillance and monitoring processes, Abbott became aware of a limited number of tip detachment events during catheter removal from the packaging. Although the packaging has not changed since the introduction of the device (2022), improper removal from the packaging can result in tip damage. While most damaged tips (26 out of 29 instances) were detected prior to catheter insertion into patients, Abbott has experienced 3 recent instances where tips have detached and remained in the patient (incidence rate of <1:100,000). In all three (3) cases there were no further adverse outcomes reported to the patients. Abbott has made changes to the packaging tray design to help mitigate tip damage during catheter removal and has recently started to deploy the new packaging.