Health Products

BD Alaris™ Pump Module and subset of Compatible Pump Infusion Sets

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: BD Alaris™ Pump Module and subset of Compatible Pump Infusion Sets
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
BD Alaris™ Pump ModuleMore than 10 numbers, contact manufacturer.8100
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.2430-0500
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.10561554
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.2465-0007
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.10010453
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.10010454
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.11532269
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.11171447
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.10013186
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.10015645
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.10062818
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.10015414
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.2426-0500
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.10010483
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.10015012
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.2452-0007
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.10015414
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.2204-0007
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.10863358
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.2202-0500
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.2202-0007
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.2432-0007
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.2430-0500
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.2403-0007
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.2426-0007
Subset Of BD Alaris™ Compatible Pump Infusion SetsMore than 10 numbers, contact manufacturer.10942011
BD Alaris™ PCUMore than 10 numbers, contact manufacturer.8015

Issue

BD is initiating this voluntary recall notice to inform customers of worse performance, under certain use cases, for the BD Alaris™ Pump Module model 8100 (pump module) when used with a subset of compatible pump infusion sets (see Appendix A). Pump infusion set design features such as in-line filters, back check valves, and Y-sites were identified to adversely impact infusion performance, particularly at programmable rates under 1 mL/h and at or under 1 mL of volume. Compared to the performance described in the user manual, the flow rate and bolus dose accuracy for a subset of compatible pump infusion sets have decreased, and the time to alarm for upstream and downstream occlusions and associated post-occlusion bolus have increased. In this letter, BD is sharing the latest performance data, reiterating existing warnings and mitigations from user manuals, and requesting customers to acknowledge and disseminate this notice within their institutions. Update (2026-09-28): BD will issue an updated recall notice as some Alaris™ pump sets with restrictive components and designs (e.g., closed male luer, filters, check valves, y-sites, microbore tubing, etc.) Perform worse than previously disclosed. The deviations most commonly occur at low programmable rates under 1 ml/h and at or under 5 ml of "bolus volume". BD has continued to extensively test BD Alaris™ pump infusion sets to ensure the most accurate set performance data is shared with our customers. This testing found that the previously claimed performance was overestimated, and, in some instances, the infusion performance is significantly worse.