Health Products

DLP Cardiac Vents & Sumps

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: DLP Cardiac Vents & Sumps
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
DLP Cardiac Vents & SumpsMore than 10 numbers, contact manufacturer.12115
DLP Cardiac Vents & SumpsMore than 10 numbers, contact manufacturer.12110
DLP Cardiac Vents & SumpsMore than 10 numbers, contact manufacturer.12113

Issue

Since March 2024, there has been an increase in complaints reported for the DLP malleable left heart vent catheter for the catheters resisting shape retention when being bent. The catheters are intended to be malleable and retain a bend in the shaft. It was determined that the primary supplier updated their specification to remove annealing process to attain tensile strength listed in ASTM1313-18.