· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Allura Xper FD20 (Product Of AlluraClarity Family) | Not applicable. | 722028 |
| Allura Xper FD 20 Or Table (Product Of AlluraClarity Family) | Not applicable. | 722035 |
| Allura Xper FD10/10 (Product Of AlluraClarity Family) | Not applicable. | 722027 |
| Allura Xper FD10 (Product Of AlluraClarity Family) | Not applicable. | 722026 |
| Allura Xper FD20/20 (Product Of AlluraClarity Family) | Not applicable. | 722038 |
| Allura Xper FD10 | Not applicable. | 001443 |
| Allura Xper FD20/15 | Not applicable. | 722058 |
Philips has identified that the BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the Allura System will be halted and the Allura System will not start. No user messages are shown before the battery is low on power or depleted.