Health Products

Allura Xper Systems

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Allura Xper Systems
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Allura Xper FD20 (Product Of AlluraClarity Family)Not applicable.722028
Allura Xper FD 20 Or Table (Product Of AlluraClarity Family)Not applicable.722035
Allura Xper FD10/10 (Product Of AlluraClarity Family)Not applicable.722027
Allura Xper FD10 (Product Of AlluraClarity Family)Not applicable.722026
Allura Xper FD20/20 (Product Of AlluraClarity Family)Not applicable.722038
Allura Xper FD10Not applicable.001443
Allura Xper FD20/15Not applicable.722058

Issue

Philips has identified that the BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the Allura System will be halted and the Allura System will not start. No user messages are shown before the battery is low on power or depleted.