Details

Alert type: Health product recall
Category: Health products - Medical devices - Orthopaedics
Product: Low Prof Lckng Screw F/Im Nail, Sterile
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Low Prof Lckng Screw F/Im Nail, Sterile04.045.346TS31756P7
Low Prof Lckng Screw F/Im Nail, Sterile04.045.342TS31756P9

Issue

The affected product lots are being recalled because, although the etched length on the screw is correct, the length marked on the external packaging is incorrect. Specifically, 42mm screws were packaged and labeled as 46mm, while 46mm screws were packaged and labeled as 42mm.