· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Beckman Coulter AU Immunochemistry System - Bicarbonate (CO2) Assay | More than 10 numbers, contact manufacturer. | OSR6237 |
| Beckman Coulter AU Immunochemistry System - Bicarbonate (CO2) Assay | More than 10 numbers, contact manufacturer. | OSR6137 |
Customer complaints were received for multiple bicarbonate reagent lots associated with false high patient results for bicarbonate. The high results are due to elevated levels of LDH in the patient samples tested. The current bicarbonate product IFU's do not advise of LDH interference with this assay.