Health Products

Spacelabs display cables for Xprezzon® or Qube Bedside Monitors

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Spacelabs display cables for Xprezzon® or Qube Bedside Monitors
Issue: Medical devices - Device compatibility, Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
DVI Cable Used With Medical Grade Flat Panel Display With Touchscreen012-089X-XX (cable)94267
DVI Cables Used With Qube Compact Monitor012-089X-XX (cable)91390
DVI Cable Used With Bedside Monitor012-089X-XX (cable)91393

Issue

While conducting 60601-1-2 testing for ongoing product development, it was found that the display cables used to connect an Xprezzon monitor(91393) or Qube compact monitor(91390) to an Xprezzon display(94267) radiates emissions outside the approved limits.