Health Products

RADspeed Pro X-Ray and CH-200, CH-200M Ceiling Tube Supports

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: RADspeed Pro X-Ray and CH-200, CH-200M Ceiling Tube Supports
Issue: Medical devices - Performance
What to do: Countermeasure parts will prevent detachment if the shaft fails.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
CH-200, CH-200M Ceiling Tube SupportMore than 10 numbers, contact manufacturer.CH-200M
CH-200, CH-200M Ceiling Tube SupportsMore than 10 numbers, contact manufacturer.CH-200
RADspeed Pro X-Ray SupportNot applicable.Not applicable.

Issue

The CH-200 and CH-200M ceiling tube supports include a suspension mechanism and rotating shaft that allow x-ray tube positioning. It was found that the mounting shaft may crack and eventually break due to material aging. in rare cases, the protective components may not function properly if the shaft fails while the unit is lifted forward.