· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Access 2 Reaction Vessels | P243971 | 81901 |
| Access 2 Reaction Vessels | P244345 | 81901 |
| Access 2 Reaction Vessels | P250005 | 81901 |
| Access 2 Reaction Vessels | P250001 | 81901 |
| Access 2 Reaction Vessels | P243943 | 81901 |
| Access 2 Reaction Vessels | P244202 | 81901 |
| Access 2 Reaction Vessels | P244344 | 81901 |
| Access 2 Reaction Vessels | P250002 | 81901 |
| Access 2 Reaction Vessels | P250004 | 81901 |
| Access 2 Reaction Vessels | P250003 | 81901 |
Instrument testing replicated failures seen in the field (e.g., pipetting, rake, wash carousel errors) for suspect RV lots. Inspections confirmed high percentages of rigid gate vestige exceeding specifications across multiple lots.