Health Products

Philips Lumify Systems

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology, Health products - Medical devices - Cardiovascular
Product: Philips Lumify Systems
Issue: Medical devices - Device compatibility
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Lumify Diagnostic Ultrasound System iOSMore than 10 numbers, contact manufacturer.795161
Lumify Power ModuleMore than 10 numbers, contact manufacturer.453561998452

Issue

Philips has become aware that the Lumify system is incompatible with Apple devices that have USB-C connector and are running iOS 18, which will cause a failure to perform live imaging potentially resulting in the procedure needing to be redone or rescheduled.