· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Coulter DxH Diluent | More than 10 numbers, contact manufacturer. | 628017 |
Beckman Coulter has confirmed that specific diluent lot numbers of Coulter DxH Diluent contribute to elevated platelet (PLT) Daily Checks background counts cycle when used with the UniCel DxH 600/690T/800/900 analyzers. This may result in elevated PLT background counts in Daily Checks ranging from 3.0 to 9.1 x 103 cells/μL.