Health Products

Benevision Central Monitoring System (CMS)

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Benevision Central Monitoring System (CMS)
Issue: Medical devices - Performance
What to do: Mindray DS USA, INC. TEL: 888-569-9225
Audience: Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Benevision Central Monitoring System (CMS)V.6.0.9484.1Not applicable.
Benevision Central Monitoring System (CMS)V. 6.0.9484.1Not applicable.

Issue

Shenzhen Mindray has identified that when the workstation number 115-050935-00 of the Benevision Central Monitoring (CMS) System has a specific hardware configuration (HP Pro Mini 400 G9 desktop PC), the computer may experience audio playback failure or screen freezing. This is caused by an issue with the 6.0.9484.1 version of the audio driver that causes excessive and abnormal system memory consumption. To date, there have been no reports of adverse events related to this issue.