Health Products

Zeus Infinity Empowered (Anesthesia Workstation)

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Zeus Infinity Empowered (Anesthesia Workstation)
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Zeus Infinity Empowered (Anesthesia Workstation)More than 10 numbers, contact manufacturer.MK04000

Issue

The action is being initiated due as a cybersecurity vulnerability was discovered as part of an internal SW penetration test. The USB interface was identified as a possible gateway. There have not been any reports of patient injuries to date.