· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue |
|---|---|---|
| Beckman Coulter AU Immunochemistry System- HDL/LDL Cholesterol Assay | More than 10 numbers, contact manufacturer. | OSR6196 |
| Beckman Coulter AU Immunochemistry System- HDL/LDL Cholesterol Assay | More than 10 numbers, contact manufacturer. | OSR6296 |
LDL cholesterol OSR6x96 does not meet the triglyceride interference claim per the BAOSR6x96 v10 IFU which states "no significant interference up to 1,500 mg/dl triglyceride" where the criteria for no significant interference is recovery within 10% of the initial value as outlined in the IFU.