Health Products

Suture/Laceration Tray by Canadian Hospital Specialties Ltd.

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Suture/Laceration Tray by Canadian Hospital Specialties Ltd.
Issue: Medical devices - Labelling and packaging
What to do: Contact the manufacturer if you require additional information.
Audience: Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Suture/Laceration Tray20892085-4361

Issue

The recall is due to a labeling error affecting the Laceration Tray MRX. The kits that manufactured with lot #208920, have discrepancy in the content and the product label. The original product contains fine teeth forceps; however, the affected lot states Adson forceps without teeth on the label. Customers should check their inventory and destroy them as per their facility process if they are part of the affected lot.