Health ProductsType I recall

Broselow™/Hinkle Pediatric Emergency System Sets

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Broselow™/Hinkle Pediatric Emergency System
Issue: Medical devices - Device compatibility
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Healthcare, Industry
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Broselow™/Hinkle Pediatric Emergency System Emergency Tape00043384397700RE
Broselow™/Hinkle Pediatric Emergency System Emergency Tape00043290787700RE
Broselow™/Hinkle Pediatric Emergency System Emergency Tape00043290797700RE
Broselow™/Hinkle Pediatric Emergency System Emergency Tape00043255837700RE
Broselow™/Hinkle Pediatric Emergency System Emergency Tape00043312557700RE
Broselow™/Hinkle Pediatric Emergency System Emergency Tape00043290777700RE
Broselow™/Hinkle Pediatric Emergency System Emergency Tape00043333687700RE
Broselow™/Hinkle Pediatric Emergency System Emergency Tape00043355377700RE
Broselow™/Hinkle Pediatric Emergency System Kit00043176997730IALS
Broselow™/Hinkle Pediatric Emergency System Kit00043335937730IALS
Broselow™/Hinkle Pediatric Emergency System Kit00043331717730IALS

Issue

AirLife™ has identified three medication-related errors in the content printed on the affected product version of the AirLife™ Broselow™ Rainbow Tape. Incorrect dosage of ketamine, flumazenil, and vecuronium is listed on the color zones of the Broselow™ tape which could potentially lead to an adverse event.