· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Broselow™/Hinkle Pediatric Emergency System Emergency Tape | 0004338439 | 7700RE |
| Broselow™/Hinkle Pediatric Emergency System Emergency Tape | 0004329078 | 7700RE |
| Broselow™/Hinkle Pediatric Emergency System Emergency Tape | 0004329079 | 7700RE |
| Broselow™/Hinkle Pediatric Emergency System Emergency Tape | 0004325583 | 7700RE |
| Broselow™/Hinkle Pediatric Emergency System Emergency Tape | 0004331255 | 7700RE |
| Broselow™/Hinkle Pediatric Emergency System Emergency Tape | 0004329077 | 7700RE |
| Broselow™/Hinkle Pediatric Emergency System Emergency Tape | 0004333368 | 7700RE |
| Broselow™/Hinkle Pediatric Emergency System Emergency Tape | 0004335537 | 7700RE |
| Broselow™/Hinkle Pediatric Emergency System Kit | 0004317699 | 7730IALS |
| Broselow™/Hinkle Pediatric Emergency System Kit | 0004333593 | 7730IALS |
| Broselow™/Hinkle Pediatric Emergency System Kit | 0004333171 | 7730IALS |
AirLife™ has identified three medication-related errors in the content printed on the affected product version of the AirLife™ Broselow™ Rainbow Tape. Incorrect dosage of ketamine, flumazenil, and vecuronium is listed on the color zones of the Broselow™ tape which could potentially lead to an adverse event.