Health Products

CellMek SPS

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Haematology
Product: CellMek SPS
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberModel or catalogue numberAffected products
SW Version: 2.3.90.0C44603CellMek SPS

Issue

Beckman Coulter has confirmed an error in SW version 2.3.90.0 that causes samples to be dispensed unexpectedly close to the bottom of the output tubes. The intended design is for samples to be dispensed above the fluid surface into the output tubes.