· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Medlite Homecare Bed | More than 10 numbers, contact manufacturer. | MDR107003LO |
| Medline Basic Homecare Bed | More than 10 numbers, contact manufacturer. | MDR107002E |
| Medline Basic Homecare Bed | More than 10 numbers, contact manufacturer. | MDR107003E |
| Medlite Homecare Bed | More than 10 numbers, contact manufacturer. | MDR107002L |
| Medlite Homecare Bed | More than 10 numbers, contact manufacturer. | MDR107003L |
| Medline Basic Homecare Bed | More than 10 numbers, contact manufacturer. | MDR107003ELO |
Medline identified the following two risks with Medline homecare beds through post-market surveillance: (1) electrical system safety risks which may result in fire and (2) the use of non-Medline accessories with the beds may have the potential for entrapment. Customers will be advised to read and follow the notice.