BS 3/8x3/32 L1.7 (Bubble Sensor for 3/8” X 3/32” tubing, length 1,7 m)

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: BS 3/8x3/32 L1.7 (Bubble Sensor for 3/8” X 3/32” tubing, length 1,7 m)
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Lot or serial numberAffected productsModel or catalogue number
More than 10 numbers, contact manufacturer.BS 3/8x3/32 L1.7 (Bubble Sensor for 3/8” X 3/32” tubing, length 1,7 m)70105.5720

Issue

Internal investigations have identified an issue with the durability of the connecting cable near the connection to the Bubble Sensor. Excessive bending of the cable can lead to poor contact, which may trigger errors on the connected medical device. These errors can occur temporarily when the cable is moved or permanently if the connection is fully compromised.