Health ProductsType I recall

Hamilton-C2 Intensive Care Ventilator

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Hamilton-C2 Intensive Care Ventilator
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type I
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Hamilton-C2 Intensive Care VentilatorAll lots.160001

Issue

Hamilton Medical AG became aware of a malfunction with HAMILTON-C2 ventilators when used long term with neonatal patients. If an affected HAMILTON-C2 ventilator is used cumulatively for 91 days with neonatal patient without a restart of the device, the device will switch to "ambient state", resulting in an immediate stop of ventilation. During the "ambient state" the device is alarming audibly and visually until it is switched off by the user. Hamilton Medical AG previously communicated this issue through FSCAs with reference CER 110972 (May 2023) and FSCA-2025-01-01 (January 2025). This updated FSN is provided with a new reference number and addresses only HAMILTON-C2 where the neonatal option is installed. Since the initiation of the first field action in May 2023, Hamilton Medical AG did not receive any new complaint related to the issue for HAMILTON-C2 devices.

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime