Health Products

VITEK® 2 Systems

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Microbiology
Product: VITEK® 2 Systems
Issue: Medical devices - Performance
What to do: BioMérieux recommends confirmatory testing of resistant results for Enterobacterales/Enterobacteriaceae and P. aeruginosa on VITEK® 2 GN-AST cards for select antibiotics.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
VITEK 2 SystemNot applicable.AST-N804
VITEK 2 SystemNot applicable.AST-XN42
VITEK 2 Compact 30 SystemsNot applicable.Not applicable.
VITEK 2 Compact 60 SystemNot applicable.Not applicable.

Issue

BioMérieux identified increased out-of-range high quality control results for the following antibiotic formulations: ceftazidime/avibactam (CZA02n), imipenem (IPM05n), imipenem/relebactam (IPR01n), meropenem/vaborbactam (MEV01n), and polymyxin B (PB02n), as well as reports of false resistance on VITEK® 2 GN-AST Cards when testing Enterobacterales/Enterobacteriaceae and P. aeruginosa.