· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| cobas® c 303 analytical unit | versions < 01-04 | 09031529001 |
| cobas® pro sample supply unit | versions < 03-01 | 08464502001 |
| cobas® pro sample supply unit | versions < 03-01 | 09205632001 |
| cobas® pure sample supply unit | versions < 01-04 | 09031537001 |
| cobas® c 503 analytical unit | versions < 03-01 | 08463662001 |
A rare software malfunction has been identified on the cobas® pro (c503) and cobas® pure (c303) platforms affecting the calibration library. This defect only occurs with cobas® pro software version <03-01 or cobas® pure software version <01-04. It can impact assays using spline-type calibration.