Health Products

cobas® pro and cobas® pure

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: cobas® pro and cobas® pure
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
cobas® c 303 analytical unitversions < 01-0409031529001
cobas® pro sample supply unitversions < 03-0108464502001
cobas® pro sample supply unitversions < 03-0109205632001
cobas® pure sample supply unitversions < 01-0409031537001
cobas® c 503 analytical unitversions < 03-0108463662001

Issue

A rare software malfunction has been identified on the cobas® pro (c503) and cobas® pure (c303) platforms affecting the calibration library. This defect only occurs with cobas® pro software version <03-01 or cobas® pure software version <01-04. It can impact assays using spline-type calibration.