Health Products

Raha Generator Shield

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Raha Generator Shield
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Raha Generator Shield50410014SWE-EP1740-001

Issue

SWE was indirectly made aware of a circumstance wherein our customer reversed the eluate/eluent lines and thereby caused a radioactive leak.