Paradigm and MiniMed 600 and 700 series pumps

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: Paradigm and MiniMed 600 and 700 series pumps
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: General public
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Minimed 630gAll lots.MMT-1714
Paradigm 754 Insulin PumpAll lots.MMT-754
Minimed 780g KitAll lots.MMT-1895
Minimed 670gAll lots.MMT-1782
Paradigm 554 Insulin PumpAll lots.MMT-554
Minimed 770gAll lots.MMT-1891
Minimed 770gAll lots.MMT-1881
Minimed 670g KitAll lots.MMT-1762
Minimed 780g PumpAll lots.MMT-1885

Issue

As part of the testing for the Alternate Controller Enabled (ACE) device submission, the R&D team tested the Minimed 780g Insulin Pump at elevated and lowered height conditions relative to the infusion site per the engineering test report. The results of the testing showed that there is an unintended over or under delivery of insulin that occurs when the 780g pump is either elevated or lowered relative to the infusion site.