Health Products

Olympus MAJ-1443 and MAJ-1444 Valves

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Gastroenterology and urology, Health products - Medical devices - General hospital and personal use
Product: Olympus MAJ-1443 and MAJ-1444 Valves
Issue: Medical devices - Labelling and packaging
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Air/Water ValvesAll lots.MAJ-1444
Reusable Suction ValveAll lots.MAJ-1443
Endoscope ReprocessorAll lots.OER-ELITE
OER-Pro Endoscope ReprocessorAll lots.OER-PRO

Issue

Olympus is informing that the MAJ-1443 and MAJ-1444 are no longer reprocessing compatible with the OERPro and OER-Elite automated endoscope reprocessors. This means the MAJ-1443 and MAJ-1444 can no longer be high-level disinfected using the OER-Pro or the OER-Elite and must instead be reprocessed by methods described in updated IFU.