· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| PKS™ Cutting Forceps | More than 10 numbers, contact manufacturer. | 920005PK |
| Everest Bipolar 5 mm Cutting Forceps | More than 10 numbers, contact manufacturer. | 3005 |
| PK® Cutting Forceps 5 mm | More than 10 numbers, contact manufacturer. | PK-CF0533 |
| HALO™ PKS™ Cutting Forceps | More than 10 numbers, contact manufacturer. | HACF0533 |
Olympus identified inadequate supplier welding validation on several 5 mm cutting forceps models, which may cause jaw breakage during clinical use. Due to this defect, Olympus is requesting customers to return all affected products to prevent potential safety risks and ensure proper device performance.