Health Products

Mixing Bowl And Spatula, 20- Pack

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Orthopaedics
Product: Mixing Bowl And Spatula, 20- Pack
Issue: Medical devices - Sterility
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Mixing Bowl And Spatula, 20- Pack8539002800-5049-011-00
Mixing Bowl And Spatula, 20- Pack8239523800-5049-011-00
Mixing Bowl And Spatula, 20- Pack8539002200-5049-011-00
Mixing Bowl And Spatula, 20- Pack8539002400-5049-011-00

Issue

Zimmer Biomet is conducting a lot specific medical device recall removal for the mixing bowl and spatula. Ten complaints have been received identifying issues at the time of use related to the package seal, including incomplete seals, wrinkles in the seals or peeling seals. Seal integrity testing conducted internally showed potential sterility breach in 7.5% of the tested units that had visual seal nonconformities. The devices are distributed in 20-pack boxes. The defectiveness was discovered on November 19, 2025. Issue was identified through the complaints handling process.