SynchroMed Clinician Programmer Application

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General hospital and personal use
Product: SynchroMed Clinician Programmer Application
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
SynchroMed Clinician Programmer ApplicationMore than 10 numbers, contact manufacturer.A810

Issue

Complaint investigations identified a software sequencing issue in the Medtronic A810 SynchroMed Clinician Programmer Application (CP app), specifically affecting flex dosing when programming the SynchroMed 8637 and 8667 pumps. When flex dosing steps are added overlapping and then out of chronological order, the app may fail to reorder them correctly, resulting in infusion steps being delivered at incorrect times. This can lead to missed flex doses, underdosing, or shifts in dose timing, and no immediate alerts. The issue is limited to flex dosing and does not impact other infusion modes. Eleven (11) complaints have been reported globally related to this issue.