· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| Ingenia 1.5T | 70851 | 1.5T |
| Ingenia Elition X | 45052 | 781358 |
| Ingenia Elition X | 45590 | 782107 |
| Ingenia Elition X | 62158 | 782107 |
| Ingenia Ambition X | 49004 | 782138 |
| Achieva 3.0T MRI System - Main Unit | 17385 | ACHIEVA 3.0T |
| Achieva 3.0T MRI System - Main Unit | 38109 | ACHIEVA 3.0T |
Philips has become aware through the MRE software system developer Resoundant about an issue affecting MRE stiffness measurements when a specific range of image reconstruction parameters is used in combination with Resoundant's algorithm. Specifically, the reconstruction voxel size settings in the default MRE scan protocol are too small, which can potentially lead to inaccurate liver stiffness measurements (lower values).