Health Products

Chuck For Pins Hoffmann3 Universal For Ø4/5/6mm

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Orthopaedics
Product: Chuck For Pins Hoffmann3 Universal For Ø4/5/6mm
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Chuck For Pins Hoffmann3 Universal For Ø4/5/6mmAR400049229050
Chuck For Pins Hoffmann3 Universal For Ø4/5/6mmAN426149229050
Chuck For Pins Hoffmann3 Universal For Ø4/5/6mmAU079449229050
Chuck For Pins Hoffmann3 Universal For Ø4/5/6mmAP376149229050
Chuck For Pins Hoffmann3 Universal For Ø4/5/6mm1544AB49229050

Issue

Stryker confirmed that certain lots of Hoffman 3 Universal Pin Chucks contained inhomogeneous coating with out-of-specification coating thickness and color. This nonconformance was discovered via product complaint.

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime