Health Products

Cook Medical Beacon® Tip Centimeter Sizing Catheter

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Cardiovascular
Product: Cook Medical Beacon® Tip Centimeter Sizing Catheter
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Beacon® Tip Centimeter Sizing CatheterMore than 20 numbers, contact manufacturer.N5.0-35-100-P-10S-0-CSC-20
Beacon® Tip Centimeter Sizing CatheterMore than 20 numbers, contact manufacturer.NR5.0-35-70-P-10S-0-CSC-20
Beacon® Tip Centimeter Sizing CatheterMore than 20 numbers, contact manufacturer.N5.0-35-100-P-10S-PIG-CSC-20

Issue

Cook Medical identified that the marker bands on impacted products may be at an increased risk of cracking/breakage. The subject marker bands are provided to Cook by an external supplier. This issue was identified internally during production. Following issue detection, product samples underwent tensile testing, where one sample cracked while being manipulated to load into the test fixture. While there is a quality control check in place to detect this failure mode prior to product release, all tested samples had underwent this quality control check prior to being tested. This is indicative that the subject marker bands may fail at a point past the quality control check. This issue has been isolated to specific supplier lots of these marker bands.

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime