· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| SILASTIC® Foley Catheter | More than 10 numbers, contact manufacturer. | 33416 |
| SILASTIC® Foley Catheter | More than 10 numbers, contact manufacturer. | 33420 |
| SILASTIC® Foley Catheter | More than 10 numbers, contact manufacturer. | 33620 |
| SILASTIC® Foley Catheter | More than 10 numbers, contact manufacturer. | 33622 |
| SILASTIC® Foley Catheter | More than 10 numbers, contact manufacturer. | 33418 |
| SILASTIC® Foley Catheter | More than 10 numbers, contact manufacturer. | 33422 |
| SILASTIC® Foley Catheter | More than 10 numbers, contact manufacturer. | 33616 |
| SILASTIC® Foley Catheter | More than 10 numbers, contact manufacturer. | 33614 |
| SILASTIC® Foley Catheter | More than 10 numbers, contact manufacturer. | 33618 |
| SILASTIC® Foley Catheter | More than 10 numbers, contact manufacturer. | 33424 |
| SILASTIC® Foley Catheter | More than 10 numbers, contact manufacturer. | 33624 |
BD has received complaints, and confirmed through internal inspection that stains are present on the surface of the affected SILASTIC® catheters. The cause of the stains was determined to be an increased amount of a raw material used in the manufacturing process. This raw material may be harmful if present in increased amounts on the surface of the catheter.