· Published by Health Canada · Official notice
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| EPOC Blood Analysis BGEM Test Cards | 01-26030-70 | 10736515 |
| EPOC Blood Analysis BGEM Test Cards | 02-26028-80 | 10736515 |
| EPOC Blood Analysis BGEM Test Cards | 02-26027-80 | 10736515 |
| EPOC Blood Analysis BGEM Test Cards | 01-26024-70 | 10736515 |
| EPOC Blood Analysis BGEM Test Cards | 01-26026-70 | 10736515 |
| EPOC Blood Analysis BGEM Test Cards | 01-26028-70 | 10736515 |
Siemens Healthineers has confirmed the occurrence of discrepant low pH and measured total carbon dioxide (MTCO2) results in arterial, venous, and capillary patient samples.