Hamilton Medical AG Breathing Set MRI

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Anaesthesiology
Product: Hamilton Medical AG Breathing Set MRI
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Model or catalogue numberLot or serial numberAffected products
260168Not applicable.Breathing Set MRI
260167Not applicable.Breathing Set MRI

Issue

Certain lots of the expiratory valve set pre-assembled to the breathing circuit set, coaxial (see FSN) do not perform as intended. In some cases, the expiratory valve membrane adhered to the expiratory valve body's sealing ring, thereby impairing valve opening and expiratory gas flow. The issue is not detected during the pre-operative test of the ventilator unless a test lung is actively ventilated. In case of an adhering expiratory valve membrane, the malfunction appears within the first breaths after ventilation is initiated. When the expiratory gas flow is impeded, the end-expiratory pressure is too high, or the end-expiratory flow is too low. Consequently, the ventilator triggers the high priority alarm "Exhalation Obstructed". The obstructed air is released through the device, e.g.the obstruction valve opens so that the patient can exhale through the inspiratory limb.

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime