ARTIS Icono Ceiling, Floor and Biplane Systems Jun 3, 2026 · Published by Health Canada · Official notice
Share
Alert type : Health product recall
Category : Health products - Medical devices - Radiology
Product : ARTIS Icono Ceiling, Floor and Biplane Systems
Issue : Medical devices - Performance
What to do : Contact the manufacturer if you require additional information.
Audience : General public, Healthcare, Industry
Recall class : Type II
We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.
Get health product recall alerts CA$3.99/mo · cancel anytime
Affected products Lot or serial number Model or catalogue number ARTIS Icono Ceiling Software version VE40A & VE40B 11328100 ARTIS Icono Floor Software version VE40A & VE40B 11327700 ARTIS Icono Biplane Software version VE40A & VE40B 11327600
In very rare situations, after increasing the detector lift height to its maximum or minimum, the system may lose its movement capability. A reboot may not reestablish movement control. This failure can be resolved only by qualified service personnel.