Health Products

Optima XR240amx X-Ray System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Radiology
Product: Optima XR240amx X-Ray System
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Optima XR240amx X-Ray System905522AMX2405737545
Optima XR240amx X-Ray System519941XR240A5737545

Issue

Mobile X-ray column assembly issue on certain Optima XR200/220/240amx systems.

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime