Health Products

RAPIDLab® 348EX System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - Chemistry
Product: RAPIDLab® 348EX System
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: General public, Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
RAPIDLab® 348EX SystemB777RAPIDLAB348EX
RAPIDLab® 348EX SystemD065RAPIDLAB348EX
RAPIDLab® 348EX SystemE457RAPIDLAB348EX

Issue

Siemens Healthcare is informing affected customers that a software-related behavior identified in the RAPIDLab® 348EX System associated with the calcium ion concentration adjusted to H 7.4 result when the instrument is connected to a Laboratory Information System (LIS).