Product: WASPLab
Issue: Medical devices - Performance
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
WASPLabW087-0147W087
WASPLabW087-0148W087

Issue

Copan WASP SRL is performing a medical device product correction related to the WASPLab product. A software bug has been identified in the webapp software which, under specific conditions, may prevent the result suggested by phenomatrix from being overwritten by the result selected by the user. As a consequence, the result transmitted by WASPLab to the LIS may be the result suggested by phenomatrix, rather than the result selectedby the user.