Health Products

Omni-Tract Surgical Retractor System

· Published by Health Canada · Official notice

Details

Alert type: Health product recall
Category: Health products - Medical devices - General and plastic surgery
Product: Omni-Tract Surgical Retractor System
Issue: Medical devices - Sterility
What to do: Contact the manufacturer if you require additional information.
Audience: Healthcare, Industry
Recall class: Type II
Get told about the next health product recall.

We email or text you when Canada issues a recall that matches what you watch: a brand, a product, an allergen. Most never make the news.

Get health product recall alerts

CA$3.99/mo · cancel anytime

Products

Affected productsLot or serial numberModel or catalogue number
Omni-Tract Surgical Retractor SystemAll lots.Not applicable

Issue

Integra Lifesciences is issuing a correction to the Integra® Omni-Tract® Surgical Retractor System - existing Instructions For Use (IFU) related to prevacuum sterilization dry time requirements, to ensure the Omni-Tract® devices and components are properly dried during sterilization to qualify current ISO standards. The current Omni-Tract® Surgical Retractor System IFU states a dry time of 30 minutes. The validated drying times utilizing standard pre-vacuum temperature and time standards were significantly longer and needed to be extended to 90 minutes, 110 minutes, or 120 minutes depending on the sterilization cycle exposure time used. Update: Integra LifeSciences is issuing this additional updated correction to the Integra® Omni-Tract® Surgical Retractor System - Instructions for Use (IFU) adding prevacuum sterilization parameters when using a rigid container. Previously, in our June 2026 communication, Integra LifeSciences provided drying times solely for polypropylene wrapped sterilization trays informing you of three (3) the dry times (90 minutes, 110 minutes, or 120 minutes) depending on the sterilization cycle exposure time. This updated notification contains additional sterilization testing validated using a rigid container with the updated prevacuum sterilization cycle parameters.

Get told about the next health product recall.
Get health product recall alerts

CA$3.99/mo · cancel anytime